User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: HP 3435A DMM -- With Data DATE: 22-Apr-87 AUTHOR: User Contributed REVISION: V1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 92 NUMBER OF LINES: 182 CONFIGURATION: Fluke 5200A/5215A Fluke 5440B CONFIGURATION: Fluke 5450A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R F W 1.002 HEAD 1.003 OPBR Do you wish to test AC VOLTAGE? 1.004 JMPF 36.003 1.005 HEAD {AC VOLTAGE VERIFICATION} 1.006 DISP Connect the UUT V-OHM and COM inputs to the 1.006 DISP ALTERNATING VOLTAGE CALIBRATOR OUTPUT terminals. 1.006 DISP Set the UUT to the AC VOLTS function, the 200mV range. 1.007 5200 200 10mV .5U 30H 2W 2.001 5200 200 10mV .3U 50H 2W 3.001 5200 200 10mV .3U 1KH 2W 4.001 5200 200 10mV .3U 10KH 2W 5.001 5200 200 10mV .3U 20KH 2W 6.001 5200 200 10mV 1.2U 50KH 2W 7.001 5200 200 10mV 1.2U 100KH 2W 8.001 5200 200 50mV 1.1U 30H 2W 9.001 5200 200 50mV .5U 50H 2W 10.001 5200 200 50mV .5U 20KH 2W 11.001 5200 200 50mV 17U 100KH 2W 12.001 5200 200 100mV 1.8U 30H 2W 13.001 5200 200 100mV .6U 50H 2W 14.001 5200 200 100mV .6U 500H 2W 15.001 5200 200 100mV .6U 20KH 2W 16.001 5200 200 100mV 2.5U 100KH 2W 17.001 DISP Set the UUT to the 2V range. 17.002 5200 2 1V .018U 30H 2W 18.001 5200 2 1V .006U 50H 2W 19.001 5200 2 1V .006U 1KH 2W 20.001 5200 2 1V .006U 20KH 2W 21.001 5200 2 1V .025U 100KH 2W 22.001 DISP Set the UUT to the 20V range. 22.002 5200 20 10V .18U 30H 2W 23.001 5200 20 10V .06U 50H 2W 24.001 5200 20 10V .06U 10KH 2W 25.001 5200 20 10V .06U 20KH 2W 26.001 5200 20 10V .25U 100KH 2W 27.001 DISP Set the UUT to the 200V range. 27.002 5200 200 100V 1.8U 30H 2W 28.001 5200 200 100V .6U 50H 2W 29.001 5200 200 100V .6U 15KH 2W 30.001 5200 200 100V .6U 20KH 2W 31.001 5200 200 100V 2.5U 100KH 2W 32.001 5200 * S 32.002 DISP Disconnect the UUT from the ALTERNATING VOLTAGE CALIB. 32.002 DISP Connect the UUT V-OHM and COM inputs to the 32.002 DISP PRECISION POWER AMPLIFIER OUTPUT terminals. 32.002 DISP Set the UUT to the 1200V range. 32.003 5200 1200 1000V 18U 30H 2W 33.001 5200 1200 1000V 6U 50H 2W 34.001 5200 1200 1000V 6U 1KH 2W 35.001 5200 1200 1000V 6U 10KH 2W 36.001 5200 * S 36.002 DISP Disconnect the UUT from the PRECISION POWER AMPLIFIER. 36.003 HEAD 36.004 OPBR Do you wish to test DC VOLTAGE? 36.005 JMPF 51.003 36.006 HEAD {DC VOLTAGE VERIFICATION} 36.007 DISP Connect the UUT V-OHM and COM inputs to the 36.007 DISP DIRECT VOLTAGE CALIBRATOR OUTPUT terminals. 36.007 DISP Set the UUT to the DC VOLTS function, the 200mV range. 36.008 5440 200 10mV .2U O 2W 37.001 5440 200 50mV .25U O 2W 38.001 5440 200 -100mV .3U O 2W 39.001 DISP Set the UUT to the 2V range. 39.002 5440 2 -.1V .001U O 2W 40.001 5440 2 .5V .002U O 2W 41.001 5440 2 1V .002U O 2W 42.001 DISP Set the UUT to the 20V range. 42.002 5440 20 -2V .01U O 2W 43.001 5440 20 -5V .02U 2W 44.001 5440 20 10V .02U 2W 45.001 DISP Set the UUT to the 200V range. 45.002 5440 200 20V .1U 2W 46.001 5440 200 -50V .2U 2W 47.001 5440 200 -100V .2U 2W 48.001 DISP Set the UUT to the 1200V range. 48.002 5440 1200 200V 1U 2W 49.001 5440 1200 -500V 2U 2W 50.001 5440 1200 1000V 2U 2W 51.001 5440 * S 51.002 DISP Disconnect the UUT from the DIRECT VOLTAGE CALIBRATOR. 51.003 HEAD 51.004 OPBR Do you wish to test CURRENT? 51.005 JMPF 78.003 51.006 HEAD {AC CURRENT VERIFICATION} 51.007 DISP Connect the UUT A and COM inputs to the 51.007 DISP MULTIFUNCTION CALIBRATOR OUTPUT terminals. 51.007 DISP Set the UUT to the AC mA function, the 200uA range. 51.008 5100 200 10uA 0.5U 50H 2W 52.001 5100 200 10uA 0.5U 5KH 2W 53.001 DISP Set the UUT to the 2mA range. 53.002 5100 2 0.1mA 0.005U 50H 2W 54.001 5100 2 0.1mA 0.005U 5KH 2W 55.001 DISP Set the UUT to the 20mA range. 55.002 5100 20 1mA 0.05U 50H 2W 56.001 5100 20 1mA 0.05U 5KH 2W 57.001 DISP Set the UUT to the 200mA range. 57.002 5100 200 20mA 0.6U 5KH 2W 58.001 5100 200 50mA 0.9U 5KH 2W 59.001 5100 200 100mA 1.3U 5KH 2W 60.001 DISP Set the UUT to the 2000mA range. 60.002 5100 2000 100mA 5U 5KH 2W 61.001 5100 2000 500mA 10U 5KH 2W 62.001 5100 2000 1000mA 16U 5KH 2W 63.001 5100 * S 63.002 HEAD {DC CURRENT VERIFICATION} 63.003 DISP Set the UUT to the DC mA function, the 200uA range. 63.004 5100 200 10uA 0.2U 2W 64.001 5100 200 50uA 0.4U 2W 65.001 5100 200 100uA 0.5U 2W 66.001 DISP Set the UUT to the 2mA range. 66.002 5100 2 0.1mA 0.002U 2W 67.001 5100 2 0.5mA 0.003U 2W 68.001 5100 2 1mA 0.005U 2W 69.001 DISP Set the UUT to the 20mA range. 69.002 5100 20 1mA 0.02U 2W 70.001 5100 20 5mA 0.04U 2W 71.001 5100 20 10mA 0.05U 2W 72.001 DISP Set the UUT to the 200mA range. 72.002 5100 200 10mA 0.2U 2W 73.001 5100 200 50mA 0.4U 2W 74.001 5100 200 100mA 0.5U 2W 75.001 DISP Set the UUT to the 2000mA range. 75.002 5100 2000 100mA 3U 2W 76.001 5100 2000 500mA 5U 2W 77.001 5100 2000 1000mA 8U 2W 78.001 5100 * S 78.002 DISP Disconnect the UUT from the MULTIFUNCTION CALIBRATOR. 78.003 HEAD 78.004 OPBR Do you wish to test RESISTANCE? 78.005 JMPF 92.003 78.006 HEAD {RESISTANCE VERIFICATION} 78.007 DISP Connect the UUT V-OHM and COM inputs to the 78.007 DISP RESISTANCE CALIBRATOR OUTPUT terminals. 78.007 DISP Set the UUT to the KOHM function, the 20 Ohm range. 78.008 5450 20 1Z .11U 2W 79.001 5450 20 10Z .15U 2W 80.001 DISP Set the UUT to the 200 Ohm range. 80.002 5450 200 10Z .2U 2W 81.001 5450 200 100Z .4U 2W 82.001 DISP Set the UUT to the 2K range. 82.002 5450 2 .19KZ .002U 2W 83.001 5450 2 1KZ .004U 2W 84.001 DISP Set the UUT to the 20K range. 84.002 5450 20 1.KZ .02U 2W 85.001 5450 20 10KZ .04U 2W 86.001 DISP Set the UUT to the 200K range. 86.002 5450 200 19KZ .2U 2W 87.001 5450 200 100KZ .4U 2W 88.001 DISP Set the UUT to the 2000K range. 88.002 5450 2000 190KZ 2U 2W 89.001 5450 2000 1000KZ 4U 2W 90.001 DISP Set the UUT to the 20M range. 90.002 5450 20 1MZ .03U 2W 91.001 5450 20 10MZ .10U 2W 92.001 5450 * S 92.002 DISP Disconnect the UUT from the RESISTANCE CALIBRATOR. 92.003 HEAD 92.004 OPBR Do you wish to repeat any section of the test? 92.005 JMPT 1.002 END